Sample Throughput |
1 to 96 |
Features |
Automatic scanning and identification of sample information+ nucleic acid extraction + PCR system setup |
Principle |
The principle of magnetic bead method, and the technology of rotational mixing |
Detectability |
192 assays can be completed per run, with 4 different systems of PCR-setup simultaneously |
Specimen Types |
Samples of plasma, serum, whole blood, swab solution, and urine, etc |
Sample Loading Channels |
44652 |
Pipetting Performance |
15μL and below: accuracy: A≤2%, repetitive CV ≤ 3.0%;<hr>15-50μL: accuracy: A≤1.5%, repetitive CV≤1.5%;<hr>50μL and above: accuracy: A≤1.0%, repetitive CV≤1.2%; |
Liquid Level Detection |
Capacitive-sensing level detection; pressure-sensing level detection |
Sample Tube Specifications |
Compatible with a wide range of sample collection tubes, 1.5mL and 2.0mL centrifuge tubes, cryopreservation tubes and sample cups |
Extraction Temperature Control Module |
2 steps: pyrolysis & elution<hr>with temperature control from 35 to 120°C |
Visual Inspection Technology |
Real-time inventory status monitoring of consumables and reagents |
PCR Reagent Chamber |
Light-proofing design; automatic cooling on start-up (4 ºC to 15 ºC refrigerator) |
PCR Supplies/Consumables |
Compatible with 0.1mL and 0.2mL 8-tube strips, and 96-well plates; |
Contamination Control Design |
(1) Separate, enclosed nucleic acid extraction space, with directional top ventilation to create an internal negative pressure system<hr>(2) Sample loading device equipped with airtight, drip-proof design<hr>(3) External droplet capture tray;<hr>(4) UV light sterilization units for the experimental and extraction cabinets;<hr>(5) Customized function: bottom ventilation of the nucleic acid extraction area available by special ventilation circuit design |
Information technology |
(1) Multiple sample barcodes are scanned one by one during sample rack loading<hr>(2) Identification of the extraction kit barcode, and automatic parsing of the extraction program information<hr>(3) Identification of PCR kit component barcode, and automatic parsing of amplification program information<hr>(4) Automatic parsing of PCR information and LIS connection, based on nucleic acid extraction data |
Operation |
Embedded 12-inch touch screen; multi-module real-time status light |
Size (W×D×H) |
1370×810×890 mm |
Weight |
220 Kg |
Power |
AC 220± 20 V, 50/60 Hz |