Pulse Vacuum Autoclave LT-PVA302 Catalog

Overview

Pulse Vacuum Autoclave LT-PVA302 provides a 360 L capacity and functions at a temperature of 139°C. It features pneumatically sealed and interlocked doors for secure operation. The rectangular body is made from S30408 stainless steel for longevity. This unit includes a PLC system with an intuitive touchscreen interface. Our Pulse Vacuum Autoclave offers real-time data printing and advanced safety alarms.

Specifications:

Volume 360 L
Pressure of Inner Chamber -0.1 to 0.3 MPa
Pressure of Jacket 0.3 MPa
Working Temperature 139°C
Designed Temperature 150°C
Working Pressure 0.25 MPa
Vacuum Leakage Rate ≤ 0.13 KPa/min
Ultimate Vacuum -97 KPa
Opening Pressure for Inner Chamber's Safety Valve 0.28 MPa
Opening Pressure for Jacket's Safety Valve 0.28 MPa
Precision of Sterilization Temperature Control 0 to 2°C
Temperature Uniformity ±1°C
Negative Pressure Pulsation Vacuum Amplitude and Times Amplitude -80 to 0KPa, Times 1 to 99
Times of Pulses Across Pressure Amplitude -80 to 80KPa, Times 1
Times of Positive Pressure Pulses 1 to 3
Tap Water Consumption 180 Kg/C
Steam Consumption 22 Kg/C
Power Consumption 2 KW
Power Supply 380V, 50Hz
Internal Dimension 610 × 610 × 1000 mm
External Dimension 1320 × 1350 × 1930 mm
Weight 850 Kg
Features:
  • Integrated drying function
  • GMP authentication interface
  • Quiet operation
  • Q245R material pressure vessel
  • Low maintenance
Applications :

Pulse Vacuum Autoclave LT-PVA302 is ideal for sterilizing medical devices, dressings, lumen devices, and liquids in sterile supply departments, operating rooms, and disinfecting centres.