Pulse Vacuum Autoclave LT-PVA309 Catalog

Overview

Pulse Vacuum Autoclave LT-PVA309 provides a 2500 L chamber capacity and allows inner pressure regulation from -0.1 to 0.3 MPa. The doors are pneumatically sealed and interlocked for added safety. It utilizes saturated steam to sterilize items with high heat and moisture. This unit offers optional remote monitoring and advanced maintenance systems. Our Pulse Vacuum Autoclave enhances sterilization with efficient drying and temperature control.

Specifications:

Volume 2500 L
Pressure of Inner Chamber -0.1 to 0.3 MPa
Pressure of Jacket 0.3 MPa
Working Temperature 139°C
Designed Temperature 150°C
Working Pressure 0.25 MPa
Vacuum Leakage Rate ≤0.13 KPa/min
Ultimate Vacuum -97 KPa
Opening Pressure for Inner Chamber's Safety Valve 0.28 MPa
Opening Pressure for Jacket's Safety Valve 0.28 MPa
Precision of Sterilization Temperature Control 0 to 2°C
Temperature Uniformity ±1°C
Negative Pressure Pulsation Vacuum Amplitude and Times Amplitude -80 to 0KPa, Times 1 to 99
Times of Pulses Across Pressure Amplitude -80 to 80KPa, Times 1
Times of Positive Pressure Pulses 1 to 3
Tap Water Consumption 1250 Kg/C
Steam Consumption 132 Kg/C
Power Consumption 6 KW
Power Supply 380V, 50Hz
Internal Dimension 900 × 1380 × 2000 mm
External Dimension 2360 × 1860 × 2230 mm
Weight 2700 Kg
Features:
  • Ring-type jacket design
  • GMP authentication interface
  • User-friendly touchscreen
  • Built-in alarm systems
  • S30408 stainless steel construction
  • Q245R material pressure vessel
Applications :

Pulse Vacuum Autoclave LT-PVA309 is used in central sterile supply departments, third-party disinfecting supply centres, and operating rooms for sterilizing medical devices, dressings, rubber items, implantable devices, and liquids.