Microplate Reader and Washer LT-MRW102
Microplate Reader and Washer LT-MRW102 is a fully automated system with 192 standard sample positions for high-throughput ELISA testing. It handles sample preparation, pipetting, incubation, washing, and absorbance reading with precision. Equipped with an integrated microplate reader and washing system for accurate and contamination-free results. Our Microplate Reader and Washer have advanced software and LIS support to enhance efficiency in clinical and research laboratories.
Specifications
Sample Racks |
Standard 192 sample positions (expandable, up to 552) |
Pipetting |
10 to 1000μl |
Dispensing System |
2 aspirating and dispensing probes |
Reagent Position |
23 reagent racks, editable |
Pipetting Range |
10 to 1000μl, 1μl stepping |
Pipetting Precision |
(100μl) CV ≤ 0.5% |
Dispensing Time |
4 minutes to whole 96-well microplates |
Washing Probes |
Two-line 8-nozzle manifolds |
Wash Containers |
Cleaning fluid (15L), waste water (25L), Buffer 1 (5L), Buffer 2 (2L), Buffer 3 (2L), Buffer 4 (2L), with liquid level-sensing, liquid shortage, and full alarm |
Waste Containers |
25L with waste full sensor |
Washing Residual |
< 2μl |
Reading Channels |
8 independent photometric channels, mono and dichromatic reading |
Absorbance |
0.0 to 3.0 OD |
Spectral Range |
400 to 700nm |
Optical Filters |
2 standard filters (450, 630nm); 2 more on request (405nm/450nm/492nm/630nm) |
Reading Accuracy |
±1% from 0.1–1.500, ±2% from 1.5–3.0 OD |
Incubators |
4 independent units |
Temperature Range |
RT to 45℃ |
Temperature Accuracy |
±0.5℃ |
Software System |
Above Windows 7 system |
LIS System |
Bi-directional, supports HL7 protocol |
Results |
Absorbance and results are reviewable by software |
Printer |
External optional |
Power Consumption |
400 Watt |
Power Supply |
AC220V±10% 50/60Hz; 110V±10% 60Hz |
Operating Temperature |
10 to 30℃ |
Humidity |
30% to 80% |
External Dimension |
1235×705×875 mm |
Dimension |
1380×860×1435 mm |
Net Weight |
146 kg |
Gross Weight |
220 kg |
Features
- Automated pipetting system
- Dual aspirating and dispensing probes
- Advanced reagent management
- Efficient microplate washing
- Integrated microplate reader
- Precise incubation and temperature control
- User-friendly software
Application
Microplate Reader and Washer LT-MRW102 is designed for high-throughput immunoassay processing, including precise pipetting, reagent dispensing, washing, incubation, and absorbance measurement, ensuring accurate and efficient diagnostic testing in clinical and research laboratories.